for Trending Cardiac Output and Stroke Volume in Cardiac Surgery Patients
by Joris van Houte, CZE
At the 2nd edition of the Point of Care Ultrasound Conference.
by Joris van Houte, CZE
At the 2nd edition of the Point of Care Ultrasound Conference.
Category
🤖
TechTranscript
00:03Well, good afternoon, everybody, and thank you very much for the invitation to speak
00:07on this conference.
00:09My name is Joris van Houten.
00:10I'm an anesthesiologist and intensivist at the Katherine Hospital in Eindhoven.
00:14And in this presentation, I will discuss with you the results of a study in which we evaluated
00:21the corrected carotid flow time as a non-invasive parameter for trendicardic output and stroke
00:26volume.
00:29And for those who are interested, this study was recently published in the Journal of Ultrasound
00:34in March 2022.
00:36And in the first part of this presentation, I will briefly introduce you to the carotid
00:40doppler ultrasound as a tool to measure carotid blood flow non-invasively.
00:45And here I will discuss how it may be beneficial as a point of care modality, how to perform
00:50a carotid doppler ultrasound exam, and the different parameters that can be derived from
00:55the pulse wave doppler signal.
00:58And then we'll focus more on the corrected carotid flow time itself.
01:01And during the presentation, corrected carotid flow time will be abbreviated as CCFT.
01:07In the second part of this presentation, we'll discuss the results of the aforementioned study.
01:12And hopefully, in the end, there will be some time left to answer any questions.
01:18Well, the interest of carotid doppler ultrasound as a potential point of care modality is quite
01:22understandable, as the common carotid artery has some favorable characteristics.
01:27And for example, the common carotid artery is lying very superficially in the neck of a
01:32patient and more central to the heart than the radial or brachial artery.
01:37Furthermore, the common carotid artery is easily accessible at the OR or at the ICU and may
01:43therefore be a very suitable vessel to assess non-invasively at the bedside.
01:48Only a standard echo is needed and which is available in most centers and therefore very
01:54cost effective.
01:55And furthermore, carotid doppler ultrasound allows functional dynamic monitoring.
02:02During a carotid doppler examination, the patient is lying supine with the head slightly rotated.
02:08And the common carotid artery can be easily visualized with a linear array transducer with a high frequency
02:14rate, as shown in the left lower image, by the way.
02:18And in the right upper corner, a short axis colored doppler image of the common carotid artery is seen.
02:24Just lateral from the carotid artery is the internal urelar vein.
02:30In the right lower corner, you can see the long axis view of the carotid artery
02:33as a nearly straight tube.
02:35And this is, however, not always the case.
02:38As in some people, the carotid artery may be more curved.
02:42And also, progressive arthrosclerosis may be present, which may distort the image
02:48and also greatly affects our doppler-derived parameters.
02:51For example, blood flow velocity increases greatly because of narrowing of this vessel.
03:00But let's go back to our unaffected vessel, in which a pulse wave doppler sample can be placed
03:05at the center of the carotid artery, in which the maximum angle between blood flow
03:10and the ultrasound beam will be kept below 60 degrees.
03:15And next, from the pulse wave doppler velocity signal, we can differentiate between two types of
03:20indices.
03:21On the one hand, it is blood flow velocity on the i-axis, on the y-axis, and blood flow
03:27time on the x-axis.
03:30And as we look at the y-axis, blood flow velocities can be measured as a maximum velocity, as marked
03:38by
03:38c in this picture.
03:39Or the velocity time integral, as marked by b.
03:43And from the carotid artery diameter, the cross-sectional area can be calculated,
03:48which, multiplied by the carotid velocity time integral, gives us the unilateral carotid blood flow.
03:55However, blood flow velocities are dependent on adequate doppler incination angles,
04:00and are therefore prone for measurement inaccuracies.
04:03And also, manual measurement of the carotid artery diameter is error-prone,
04:07as the diameter changes from the systolic to diastolic phase,
04:11and there is also the risk of oblique transaction of the vessel.
04:17In contrast, carotid flow time can be measured on the x-axis,
04:22and is much less dependent on incination angles,
04:24as imperfect waveforms still yield accurate flow time readings.
04:30Therefore, less training may be required for novisonogravus.
04:33The systolic flow time is the time from the foot of the systolic upstroke,
04:38until the incisera of the dichrotic notch, as depicted by b.
04:42And the complete cycle time runs from the foot of the systolic upstroke,
04:47until the foot of the next systolic upstroke, as depicted by a.
04:52In order to allow standardization, it is needed to correct the systolic flow time
04:57for heart rate, as the cycle time shortens when heart rate increases, and also vice versa.
05:04Accordingly, the corrected carotid flow time is calculated
05:07by dividing the systolic flow time by the square root of the cycle time.
05:16This carotid flow time can be seen as a proxy for left ventricular ejection time,
05:21and is therefore proportional to cardiac contractility and left ventricular preload.
05:26And it is also inversely proportional to afterload.
05:30Therefore, corrected carotid flow time may be related to a patient's cardiac output and stroke volume.
05:40In previous studies, however, up to nine different carotid corrected flow time equations have been used
05:45to calculate CCFT, of which these three equations that are mentioned here have been cited mostly.
05:52And in order to compare studies, it is relevant to know to what extent these equations lead to comparable CCFT
05:59outcomes.
06:00And this is important as CCFT values have been evaluated to determine cutoff points for fluid responsiveness.
06:09The aim of this study was therefore twofold.
06:13And in the first place, we wanted to evaluate the dynamic relationship between CCFT and invasively derived cardiac output and
06:21stroke volume.
06:21And secondly, we wanted to compare the three most commonly used CCFT equations.
06:30Therefore, we performed a prospective observational study at the OR, 18 coronary artery bypass grafting patients were included.
06:38And next, standardized anesthesia was performed, and patients were mechanically ventilated at eight mils per kilogram tidal volume and five
06:49centimeters of PEEP.
06:50Patients were excluded if they had an impaired left ventricular function, moderate to severe vulvular disease, carotid artery stenosis, previous
06:59brain injury, and severe lung disease.
07:02A Philips echo device was used, together with a Philips broadband linear area transducer.
07:08And for cardiac output measurement, the invasive and calibrated pulse contour cardiac output device, PICO, was used.
07:18Simultaneous PICO and Pulse Wave Doppler measurements from the carotid artery were obtained at three time points.
07:24Fifty minutes after the induction of anesthesia was called T1.
07:29After a passive leg raise, we repeated those measurements, which was called T2.
07:34And at the end of surgery, when patients were weaned from the extracorporeal circulation, we called that time point T3.
07:41And then six averaged Pulse Wave Doppler waveforms were manually traced offline.
07:49The patient's characteristics are shown in this table, and importantly, none of these patients had abnormal left or right ventricular
07:56function or severe vulvular disease.
07:59Therefore, the following results are not generalizable to patients with abnormal heart function.
08:09In this table, the hemodynamic parameters are shown for the three time points.
08:14So, again, T1 after induction, T2 after a passive leg raise, and T3 post-bypass.
08:21And most of these, like mean arterial pressure, stroke volume, resistive index, and tidal CO2, all remained constant.
08:29And also, anesthetic dosing was stable throughout the study protocol.
08:34At T3, cardiac output, as you can see, increased significantly.
08:39You can see that here.
08:44And that increase is most likely explained by a higher heart rate, as stroke volume remains relatively constant.
08:52Now, in this figure, the distribution of all CCFT values from T1 to T3 is shown in a violin plot.
09:00For the three equations, the width of the kernel corresponds to the numerical density of data at different CCFT values.
09:09Paired CCFT values differ significantly depending on the equation used.
09:14You can also observe this by the slight differences of the shape of the different kernels.
09:24In this figure, the correlation between cardiac output and CCFT in the upper rank, and between stroke volume and CCFT
09:30in the lower rank, can be seen.
09:33Here we observed that the correlations differed depending on the equation that was used.
09:39Sperman's correlation between CCFT and cardiac output was highest for Bazet's equation, and this correlation was moderate, with an R
09:47-value of 0.43.
09:49Sperman's correlation between CCFT and stroke volume was then highest for Woody's equation, but only weak, with an R-value
09:57of 0.33.
10:01However, the correlation coefficient may not be the best metric to assess in a dynamic study model.
10:07After all, at a certain cardiac output level, static CCFT values may differ between individuals, and therefore correlation may never
10:16be likely to be strong.
10:18Instead, the dynamic changes of hemodynamic indices in time enable evaluation of trending, which can be assessed through concordance analysis.
10:29This was shown in the next figure, and what you can see here is the delta CCFT from T1 to
10:35T3 plotted against the delta cardiac output.
10:38So, if CCFT and cardiac output either increase or decrease in the same direction, then data points are in the
10:47upper right or lower left quadrant.
10:50And then the percentage of data points in these two quadrants determine the level of concordance.
10:56If this level is less than 84%, then the concordance is only poor.
11:03If those data points are within those two quadrants, within 84% and 92%, concordance is intermediate.
11:11And above 92%, concordance is acceptable.
11:14When we look at concordance between CCFT and cardiac output in the upper rank, it was acceptable for Bezat's equation,
11:21as it was 100%.
11:24But it was only intermediate acceptable when analyzed by Chambers, and only poor when analyzed by Woody's equation.
11:32Concordance between CCFT and stroke volume, as can be seen at the lower rank, was poor for all equations, with
11:39a maximum of 82% at Woody's equation.
11:43These concordance plots were constructed with exclusion zones of 15% to eliminate low predictive data that lie near the
11:52center of the plot.
11:56In conclusion, CDU may be a promising future non-invasive human dynamic point of care modality.
12:03And compared to velocity measurements, flow time measurements are less error prone and easier to learn.
12:10Three commonly used CCFT equations lead to different CCFT values, different correlation with stroke volume and cardiac output, and different
12:20concordance rates.
12:23Now, this information is relevant when studies using different CCFT equations are being compared.
12:29The dynamic trending ability of CCFT to detect changes in cardiac output was highest, 100%, for Bezat's equation.
12:38However, the ability of CCFT to detect changes in stroke volume was only poor.
12:45Thank you for the nice presentation and applause for Joris.
12:48Thank you so much.
12:49Thank you very much.