for Trending Cardiac Output and Stroke Volume in Cardiac Surgery Patients
by Joris van Houte, CZE

At the 2nd edition of the Point of Care Ultrasound Conference.

Category

🤖
Tech
Transcript
00:03Well, good afternoon, everybody, and thank you very much for the invitation to speak
00:07on this conference.
00:09My name is Joris van Houten.
00:10I'm an anesthesiologist and intensivist at the Katherine Hospital in Eindhoven.
00:14And in this presentation, I will discuss with you the results of a study in which we evaluated
00:21the corrected carotid flow time as a non-invasive parameter for trendicardic output and stroke
00:26volume.
00:29And for those who are interested, this study was recently published in the Journal of Ultrasound
00:34in March 2022.
00:36And in the first part of this presentation, I will briefly introduce you to the carotid
00:40doppler ultrasound as a tool to measure carotid blood flow non-invasively.
00:45And here I will discuss how it may be beneficial as a point of care modality, how to perform
00:50a carotid doppler ultrasound exam, and the different parameters that can be derived from
00:55the pulse wave doppler signal.
00:58And then we'll focus more on the corrected carotid flow time itself.
01:01And during the presentation, corrected carotid flow time will be abbreviated as CCFT.
01:07In the second part of this presentation, we'll discuss the results of the aforementioned study.
01:12And hopefully, in the end, there will be some time left to answer any questions.
01:18Well, the interest of carotid doppler ultrasound as a potential point of care modality is quite
01:22understandable, as the common carotid artery has some favorable characteristics.
01:27And for example, the common carotid artery is lying very superficially in the neck of a
01:32patient and more central to the heart than the radial or brachial artery.
01:37Furthermore, the common carotid artery is easily accessible at the OR or at the ICU and may
01:43therefore be a very suitable vessel to assess non-invasively at the bedside.
01:48Only a standard echo is needed and which is available in most centers and therefore very
01:54cost effective.
01:55And furthermore, carotid doppler ultrasound allows functional dynamic monitoring.
02:02During a carotid doppler examination, the patient is lying supine with the head slightly rotated.
02:08And the common carotid artery can be easily visualized with a linear array transducer with a high frequency
02:14rate, as shown in the left lower image, by the way.
02:18And in the right upper corner, a short axis colored doppler image of the common carotid artery is seen.
02:24Just lateral from the carotid artery is the internal urelar vein.
02:30In the right lower corner, you can see the long axis view of the carotid artery
02:33as a nearly straight tube.
02:35And this is, however, not always the case.
02:38As in some people, the carotid artery may be more curved.
02:42And also, progressive arthrosclerosis may be present, which may distort the image
02:48and also greatly affects our doppler-derived parameters.
02:51For example, blood flow velocity increases greatly because of narrowing of this vessel.
03:00But let's go back to our unaffected vessel, in which a pulse wave doppler sample can be placed
03:05at the center of the carotid artery, in which the maximum angle between blood flow
03:10and the ultrasound beam will be kept below 60 degrees.
03:15And next, from the pulse wave doppler velocity signal, we can differentiate between two types of
03:20indices.
03:21On the one hand, it is blood flow velocity on the i-axis, on the y-axis, and blood flow
03:27time on the x-axis.
03:30And as we look at the y-axis, blood flow velocities can be measured as a maximum velocity, as marked
03:38by
03:38c in this picture.
03:39Or the velocity time integral, as marked by b.
03:43And from the carotid artery diameter, the cross-sectional area can be calculated,
03:48which, multiplied by the carotid velocity time integral, gives us the unilateral carotid blood flow.
03:55However, blood flow velocities are dependent on adequate doppler incination angles,
04:00and are therefore prone for measurement inaccuracies.
04:03And also, manual measurement of the carotid artery diameter is error-prone,
04:07as the diameter changes from the systolic to diastolic phase,
04:11and there is also the risk of oblique transaction of the vessel.
04:17In contrast, carotid flow time can be measured on the x-axis,
04:22and is much less dependent on incination angles,
04:24as imperfect waveforms still yield accurate flow time readings.
04:30Therefore, less training may be required for novisonogravus.
04:33The systolic flow time is the time from the foot of the systolic upstroke,
04:38until the incisera of the dichrotic notch, as depicted by b.
04:42And the complete cycle time runs from the foot of the systolic upstroke,
04:47until the foot of the next systolic upstroke, as depicted by a.
04:52In order to allow standardization, it is needed to correct the systolic flow time
04:57for heart rate, as the cycle time shortens when heart rate increases, and also vice versa.
05:04Accordingly, the corrected carotid flow time is calculated
05:07by dividing the systolic flow time by the square root of the cycle time.
05:16This carotid flow time can be seen as a proxy for left ventricular ejection time,
05:21and is therefore proportional to cardiac contractility and left ventricular preload.
05:26And it is also inversely proportional to afterload.
05:30Therefore, corrected carotid flow time may be related to a patient's cardiac output and stroke volume.
05:40In previous studies, however, up to nine different carotid corrected flow time equations have been used
05:45to calculate CCFT, of which these three equations that are mentioned here have been cited mostly.
05:52And in order to compare studies, it is relevant to know to what extent these equations lead to comparable CCFT
05:59outcomes.
06:00And this is important as CCFT values have been evaluated to determine cutoff points for fluid responsiveness.
06:09The aim of this study was therefore twofold.
06:13And in the first place, we wanted to evaluate the dynamic relationship between CCFT and invasively derived cardiac output and
06:21stroke volume.
06:21And secondly, we wanted to compare the three most commonly used CCFT equations.
06:30Therefore, we performed a prospective observational study at the OR, 18 coronary artery bypass grafting patients were included.
06:38And next, standardized anesthesia was performed, and patients were mechanically ventilated at eight mils per kilogram tidal volume and five
06:49centimeters of PEEP.
06:50Patients were excluded if they had an impaired left ventricular function, moderate to severe vulvular disease, carotid artery stenosis, previous
06:59brain injury, and severe lung disease.
07:02A Philips echo device was used, together with a Philips broadband linear area transducer.
07:08And for cardiac output measurement, the invasive and calibrated pulse contour cardiac output device, PICO, was used.
07:18Simultaneous PICO and Pulse Wave Doppler measurements from the carotid artery were obtained at three time points.
07:24Fifty minutes after the induction of anesthesia was called T1.
07:29After a passive leg raise, we repeated those measurements, which was called T2.
07:34And at the end of surgery, when patients were weaned from the extracorporeal circulation, we called that time point T3.
07:41And then six averaged Pulse Wave Doppler waveforms were manually traced offline.
07:49The patient's characteristics are shown in this table, and importantly, none of these patients had abnormal left or right ventricular
07:56function or severe vulvular disease.
07:59Therefore, the following results are not generalizable to patients with abnormal heart function.
08:09In this table, the hemodynamic parameters are shown for the three time points.
08:14So, again, T1 after induction, T2 after a passive leg raise, and T3 post-bypass.
08:21And most of these, like mean arterial pressure, stroke volume, resistive index, and tidal CO2, all remained constant.
08:29And also, anesthetic dosing was stable throughout the study protocol.
08:34At T3, cardiac output, as you can see, increased significantly.
08:39You can see that here.
08:44And that increase is most likely explained by a higher heart rate, as stroke volume remains relatively constant.
08:52Now, in this figure, the distribution of all CCFT values from T1 to T3 is shown in a violin plot.
09:00For the three equations, the width of the kernel corresponds to the numerical density of data at different CCFT values.
09:09Paired CCFT values differ significantly depending on the equation used.
09:14You can also observe this by the slight differences of the shape of the different kernels.
09:24In this figure, the correlation between cardiac output and CCFT in the upper rank, and between stroke volume and CCFT
09:30in the lower rank, can be seen.
09:33Here we observed that the correlations differed depending on the equation that was used.
09:39Sperman's correlation between CCFT and cardiac output was highest for Bazet's equation, and this correlation was moderate, with an R
09:47-value of 0.43.
09:49Sperman's correlation between CCFT and stroke volume was then highest for Woody's equation, but only weak, with an R-value
09:57of 0.33.
10:01However, the correlation coefficient may not be the best metric to assess in a dynamic study model.
10:07After all, at a certain cardiac output level, static CCFT values may differ between individuals, and therefore correlation may never
10:16be likely to be strong.
10:18Instead, the dynamic changes of hemodynamic indices in time enable evaluation of trending, which can be assessed through concordance analysis.
10:29This was shown in the next figure, and what you can see here is the delta CCFT from T1 to
10:35T3 plotted against the delta cardiac output.
10:38So, if CCFT and cardiac output either increase or decrease in the same direction, then data points are in the
10:47upper right or lower left quadrant.
10:50And then the percentage of data points in these two quadrants determine the level of concordance.
10:56If this level is less than 84%, then the concordance is only poor.
11:03If those data points are within those two quadrants, within 84% and 92%, concordance is intermediate.
11:11And above 92%, concordance is acceptable.
11:14When we look at concordance between CCFT and cardiac output in the upper rank, it was acceptable for Bezat's equation,
11:21as it was 100%.
11:24But it was only intermediate acceptable when analyzed by Chambers, and only poor when analyzed by Woody's equation.
11:32Concordance between CCFT and stroke volume, as can be seen at the lower rank, was poor for all equations, with
11:39a maximum of 82% at Woody's equation.
11:43These concordance plots were constructed with exclusion zones of 15% to eliminate low predictive data that lie near the
11:52center of the plot.
11:56In conclusion, CDU may be a promising future non-invasive human dynamic point of care modality.
12:03And compared to velocity measurements, flow time measurements are less error prone and easier to learn.
12:10Three commonly used CCFT equations lead to different CCFT values, different correlation with stroke volume and cardiac output, and different
12:20concordance rates.
12:23Now, this information is relevant when studies using different CCFT equations are being compared.
12:29The dynamic trending ability of CCFT to detect changes in cardiac output was highest, 100%, for Bezat's equation.
12:38However, the ability of CCFT to detect changes in stroke volume was only poor.
12:45Thank you for the nice presentation and applause for Joris.
12:48Thank you so much.
12:49Thank you very much.